Casella Gravimetric Adaptors - for the Microdust pro particulate monitor

Overview of some of the more popular air sampling accessories
  • Total suspended particulate (TSP)
  • sometimes called the gravimetric adaptor
  • this adaptor is used to measure all the particulate suspended in the air
  • a cassette and filter can be used to collect a simultaneous sample
  • no size selection is carried out in this method
  • an external sampling pump (such as Apex std) is required to pull the air through the probe and the adaptor
  • adaptor is used with the probe in the horizontal mode and the tube pointing downwards
click on the image for a larger view

Part # 103214B
  • Respirable adaptor
  • this adaptor incorporates a cyclone size selective system to limit the particle size that is measured
  • only particles defined as "respirable range" of 4 micron and less are collected when the flow rate is set at 2.2 l/min
  • particles can be collected in a 25 or 37 mm cassette
  • an external sampling pump (such as Apex std) is required to pull the air through the probe and the adaptor
click on the image for a larger view

Part # 103182B
  • Size selective adaptor
  • the size selective sampling adaptor uses polyurethane foam filters (PUF) to limit the size passed through the system
  • PUF filters for 10, 4 and 2.5 microns are included in the kit
  • tweezers should be used to handle the PUF filters to prevent errors
  • flow rate used with the external pump should be at 3.5 l/min
  • particles can be collected in a 25 or 37 mm cassette
  • an external sampling pump (such as Apex std) is required to pull the air through the probe and the adaptor
click on the image for a larger view

Part # 151280B adaptor kit
#151280B/K1 for kit with PUF filters
  • Aspirated adaptor
  • this adaptor has no size selection capabilities
  • all the total particulate sizes are drawn through the adaptor
  • it does not need an external pump since it has its own means of drawing the air through using a small motor to power a fan
  • the motor is driven from an external mains power supply at 110 Vac to provide a 5 Vdc voltage
  • alternatively this may be provided from a suitable small dc battery pack
  • the aspirator unit should be cleaned periodically to prevent the buildup of dust
click on the image for a larger view

Part # 103187B
includes plug top transformer
  • Laboratory adaptor
  • this adaptor can be used in long term process control application where the particulate concentration must be monitored continuously
  • no size selection is provided
  • an active system must be used to draw the air sample through the adaptor
  • can be used for studies of toxicology, smoke, air filter efficiency, etc.
click on the image for a larger view

Part # 176153B

 

  • PUF filters
  • this adaptor is based on the conical inhalable sampler design identified in the UK H&SE publication MDHS14
  • consists of polyurethane foam filters with different porosity and size
  • enable the different fractions to be separated out to get simultaneous measurements of two concentrations at a time
  • can be used to obtain the PM10 and PM2.5 fractions when used for environmental surveys
  • can be used with 4 micron filters for respirable measurements
  • can be used with the Microdust pro for outdoor measurements
  • tweezers should be used to handle the PUF filters to prevent errors
  • keep the PUF filters in a clean dust free environment
click on the image for a larger view

Part # 118200B/K1 kit with
CIS head and PUF filters,
P118204 2.5 micron,
P118206 10 micron,
P118208 4 micron
PUF filter packs
(10 PUF filters per pack)
  • DryCal/Defender 500 Series
  • use this accessory as a primary source of calibration to check the flow rate selected in the adaptor and pump
  • wide range of flow rates available for low flow and normal ranges
  • 50 mL/ m to 5Ll/m flow rate range
  • clean operation no bubbles or water used
  • very convenient and easy to use
  • quickly calibrate the flow and make adjustments to the pump if necessary
  • 510 standard range and 520 range with extra features

view product data sheet

click on the image for a larger view

Part # Bios 520 M Calibrator
  • Rotameter
  • this accessory is used to calibrate the flow rate of the pump used with the adaptor and pump
  • read the flow rate from the float in the vertical tube
  • the rotameter is a lightweight device, self powered and easy to use
  • at the end of the calibration simply dismantle and store for next use
  • can be used easily and quickly in the field even by inexperienced users
  • should be used to check the flow rate of the pump before (and after) every measurement run to ensure the pump flow has not drifted
click on the image for a larger view

Part # -CM344 Rotameter
Part # 100890B Flow meter stand

page last updated - Friday March 28, 2008